QACode | QAName | ANI | Sample results for structural isomers may have contribution from their isomeric pair |
AWM | Detection limit increased due to dilution prior to final sample volume (not a secondary dilution) |
AY | Matrix interference suspected |
BB | Sample > 4x spike concentration |
BBM | Sample > 2X but less than 4X spike concentration |
BCI | The reporting limit is elevated due to a chromatographic interference |
BCO | Sample received in improper container |
BDL | 0 < quantification (threshold) cycle (Cq) <= Cq Limit of Detection (LOD) |
BE | Low surrogate recovery; analyzed twice |
BFRE | Sample influenced by fire |
BH | Reporting limits raised due to high level of non-target analytes |
BITR | Insufficient taxonomic resolution to calculate CSCI |
BLM | Compound unidentified or below the RL due to overdilution |
BLNO | Low number of organisms, may impact CSCI calculation |
BQC | Taxonomy External Quality Control |
BRK | Sample not analyzed, sample container broken |
BRKA | Sample container broken but analyzed |
BS | Insufficient sample available to follow standard QC procedures |
BSAL | Sample influenced by saline conditions |
BSCR | Sample influenced by scour or suspected scour event |
BT | Insufficient sample to perform the analysis |
BUCM | Untested or unknown benthic macroinvertebrate collection method |
BUTP | Sampling conducted during unrepresentative time period for CSCI calculation |
BV | Sample received after holding time expired |
BWIM | Portion of upstream tributary watershed for benthic macroinvertebrates is in Mexico |
BX | Sample stored at improper temperature |
BY | Sample received at improper temperature |
BZ | Sample preserved improperly |
BZ15 | Sample not filtered within 15 minutes |
C | Pesticide result confirmed using GC/MS- EPA Flag |
CE | Sample not homogeneous |
CIN | Value outside confidence interval |
CJ | Analyte concentration is in excess of the instrument calibration; considered estimated |
CK | Initial analysis within holding time but failed QA/QC criteria |
CN | Hydrocarbon response in diesel range but does not resemble diesel |
CNC | Calibration nonconformances |
CNM | DRO (C10-C28) resembles weathered diesel |
CNP | calibration not performed |
CQA | Concentration not reported for QA sample only % Recovery or RPD reported |
CS | QC criteria not met due to analyte concentration near RL |
CSG | Bacteria result calculated using a subset of grids when the filter is too concentrated |
CT | QC criteria not met due to high level of analyte concentration |
CVH | Continuing calibration verification high, no bias |
CVHB | Continuing calibration verification high, high bias |
CZM | Project sample was heterogeneous and sample homogeneity could not be readily achieved using routine laboratory practices, therefore RPD was outside the specified acceptance limits |
D | EPA Flag - Analytes analyzed at a secondary dilution |
DB | QA results outside of acceptance limits due to matrix effects |
DBLOD | Detected Below Limit of Detection used for PCR analysis only |
DBM | QC results outside of acceptance limits, associated sample result not affected since analyte was not detected or below the reporting limit. |
DF | Reporting limits elevated due to matrix interferences |
DG | Reporting limits elevated due to sample dilution |
DO | Coelution |
DRM | Spike amount less than 5X the MDL |
DS | Batch Quality Assurance data from another project |
DT | Method detection limit elevated |
DV | Reporting limit elevated due to final volume of the sample extract was higher than the nominal amount |
E | EPA Flag - Analyte exceeded the concentration range of the GC/MS |
ERV | Exceeds reference control limits |
EUM | LCS is outside of control limits |
EX | Matrix spike diluted to not detectable during analysis |
F | Sample mistakenly filtered |
FCL | Field calibration not performed within 24 hours before use |
FDC | Drift check not acceptable |
FDI | Field instrument displaced but still deployed in water |
FDO | Field instrument displaced and removed from water |
FDP | Field duplicate RPD above QC limit |
FDPM | Field duplicate RPD above QC limit with results >=5 xRL |
FDR | Dry Channel |
FDS | Dry Site |
FEU | Equipment Unavailable |
FI | Analyte in field sample and associated blank |
FIA | Location was inaccessible to obtain a measurement |
FIB | Field instrument buried |
FIF | Instrument/Probe Failure |
FIO | Field instrument above water level |
FIP | Field instrument vandalized, fouled, or otherwise in poor condition |
FIT | Invalid time. Likely instrument error or failure |
FLV | Velocity too low to be measured |
FNM | no documentation of the field measurement collection exists |
FO | Estimated maximum possible concentration (EMPC) |
FS | Too Shallow for probe measurement |
FTD | Location was too deep to obtain a measurement |
FTT | Water too turbid to measure |
FUD | Unable to deploy instrument |
FX | Analyte present in the instrument blank |
GB | Matrix spike recovery not within control limits |
GBC | CRM analyte recovery not within control limits |
GC | Second column confirmation performed |
GCA | Second column confirmation absent (not reported) |
GD | secondary column result |
GIDA | Isotope Dilution Analogue recovery not within control limits |
GN | Surrogate recovery is outside of control limits |
GR | Internal standard recovery is outside method recovery limit |
H | A holding time violation has occurred. |
H22 | Sample exceeds incubation period of 18-22 hours. |
H24 | Holding time was > 24 hours for Bacteria tests only |
H8 | Holding time was > 8 hrs but < 24 hours for Bacteria tests only |
HB | Result positively biased |
HD | Chromat. profile inconsistent with pattern(s) of ref. fuel stnds. |
HH | Result exceeds linear range; concentration may be understated |
HL | Analyte recovery above established limit |
HM | Analyte recovery below established limit |
HNO2 | Holding time violation occurred on the NO2 concentration used for NO3 calculation. |
HR | Post-digestion spike |
HS | Spike analyte recovery is outside stated control limits |
HSM | Sampling crew or submitting agency field spike recovery not within control limits |
HT | Analytical value calculated using results from associated tests |
HX | Spike recovery out of limits, but QC batch considered in control |
IDA | Isotope Dilution Analogue corrected |
IE | Surrogate not added to this sample |
IF | Sample result is greater than reported value |
IL | RPD exceeds laboratory control limit |
ILM | RSD exceeds control limit |
ILN | Rlog exceeds control limit |
ILO | Rlog not calculated because concentration below RL |
IM | Method does not include this analyte as part of compound list |
IP | Analyte detected in field or lab generated blank |
IP10 | Analyte detected in field or lab generated blank, >RL and >1/10 amount detected in associated environmental sample |
IP5 | Analyte detected in field or lab generated blank, >RL and >1/5 amount detected in associated environmental sample |
IPF | Analyte detected in field blank, data validation code |
IPMDL2 | Analyte detected in field or lab generated blank, >10% lowest detection in batch and >2.2x MDL |
IPMDL3 | Analyte detected in field or lab generated blank, >MDL and >30% lowest detection in batch |
IPRL | Analyte detected in field or lab generated blank, >RL |
IS | Reporting limit elevated due to pres. of analyte in method blank |
IU | Percent Recovery exceeds laboratory control limit |
IZM | Bottle integrity compromised, potential contamination |
J | Estimated value - EPA Flag |
JA | Analyte positively identified but quantitation is an estimate |
JDL | Estimated result lower than detection limit |
LA | Confirmatory analysis was past holding time |
LC | Laboratory Contamination |
LRGN | Data rejected - Surrogate recovery not within control limits, flagged by laboratory |
LRIL | Data rejected - RPD exceeds laboratory control limit, flagged by laboratory |
LRIP | Data rejected - Analyte detected in field or lab generated blank, flagged by laboratory |
LRIU | Data rejected - Percent Recovery exceeds laboratory control limit, flagged by laboratory |
LRJ | Data rejected - Estimated value - EPA Flag, flagged by laboratory |
LRJA | Data rejected - Analyte positively identified but quantitation is an estimate, flagged by laboratory |
LRM | Data rejected - A matrix effect is present, flagged by laboratory |
LRQ | Data rejected - Based on professional judgement QA/QC protocols were not met, flagged by lab |
LST | Sample was lost or destroyed |
M | A matrix effect is present |
MN | Method procedures not followed |
N | Tentatively Identified Compound |
NAS | Not adjusted for salinity |
NBC | Value not blank corrected |
NC | Analyte concentration not certifiable in Certified Reference Material |
NG | The reported result is a negative value |
NMDL | No Method Detection Limit reported from laboratory |
None | None - No QA Qualifier |
NR | Not Recorded |
NRL | No Reporting Limit reported by the laboratory |
OA | Analyte over range when run undiluted. Diluted result below PQL |
OL | Result may be biased high due to carry over from prior analysis |
OV | Result is averaged |
P | Evidence analyte present |
PEH | Sample was preserved to extend the holding time |
PG | Calibration verification outside control limits |
PI | Primary and confirmation results varied by > than 40% |
PIL | RPD is within the lab or method defined acceptance criteria, but is outside of project-specific control limits |
PJ | Result from re-extract/re-anal to confirm original MS/MSD result |
PJM | Result from re-extract/re-anal to confirm original result |
PJN | Result from re-extract/re-analysis for QC purposes only |
PP | Dilutions for BOD failed criteria of 2 mg/L diss. O2 depletion |
PRM | Low survival in toxicity test resulted from test interference due to pathogen-related mortality |
Q | Questionable result |
QAX | When the native sample for the MS/MSD or DUP is not included in the batch reported |
QG | Individual analyte exceedances allowed per NELAC |
R | Data rejected - EPA Flag |
RE | Elevated reporting limits due to limited sample volume |
RLST | Replicate lost or destroyed |
ROQ | Quantification (threshold) cycle (Cq) > Cq Lower Limit of Quantification (LLOQ), positive result |
RQ | Spike recovery not calculated due to high concentration of analyte |
RQM | Spike recovery not calculated due to matrix interference |
RU | LCS rec. within marginal exceedances, 3-4 s.d. around the mean. |
RUM | Acceptable QC with marginal exceedance |
RY | Standard Instrument related QC exceeds the control limits |
RYM | Reporting Limit check standard failed laboratory defined acceptance criteria. Data is accepted based on passing method required QC |
S | Sample Receipt nonconformance |
SC | Surrogate Corrected Value |
SCR | Screening level analysis |
SLM | Analyte results for R1 and/or R2 were lower than 10 times the MDL, therefore RPD acceptance limits do not apply |
SN | See narrative and/or special notes |
SSM | Second source recovery below method control limit |
TA | Ammonia precision or accuracy exceeds laboratory control limit |
TAC | Alternative control used in toxicity statistical analysis |
TAD | Additional metamorphosized or pupated organism accidently included in statistical analysis |
TAE | Test organisms exceed age limit at test initiation |
TAF | Test organisms exceed the maximum weight requirement at test initiation |
TAS | Alternate species tested |
TC | Conductivity precision or accuracy exceeds laboratory control limit |
TCF | Alternative control does not meet test acceptability criteria |
TCI | Conductivity insufficient for test species |
TCT | Conductivity tolerance exceeded for test species |
TD | DO precision or accuracy exceeds laboratory control limit |
TEM | Endpoint not analyzed or reported by the laboratory |
TH | Hardness precision or accuracy exceeds laboratory control limit |
THS | Hydrogen sulfide precision or accuracy exceeds laboratory control limit |
TK | Alkalinity precision or accuracy exceeds laboratory control limit |
TL | Minor deviation in temperature or light (comment) |
TMM | Male replicate excluded from test analysis |
TMO | Test organisms escaped or are otherwise missing |
TMSD | Endpoint considered not toxic; per EPA method, when both the relative difference from control and the test percent minimum significant difference (PMSD) are less than EPA lower PMSD bound (10th percentile) |
TNC | Too numerous to count |
TNS | Only neonate counts from surviving adults were used in the analysis |
TOQ | Number of organisms in a toxicity test do not meet the minimum quantity per replicate at test initiation or an unequal quantity of organisms per replicate is used |
TP | pH precision or accuracy exceeds laboratory control limit |
TR | Test conditions not acceptable (temp, light) |
TRT | Reference toxicant test failed due to control survival less than 80% |
TS | Salinity precision or accuracy exceeds laboratory control limit |
TW | Water quality parameters outside recommended test method ranges |
TWN | Required water quality parameters not measured |
TXP | Toxaphene results are based on a commercial Toxaphene mixture of unknown composition and therefore the concentrations listed in this report are estimated |
UF | Counts outside acceptable range. |
UJ | Analyte was not detected above the reported sample quantitation limit.Reported quantitation limit is approx. & may not represent the actual limit of quantitation necessary to accurately and precisely measure the analyte in the sample. |
UKM | Lock mass interference present |
ULM | Analyte detected in associated bottle blank check |
UOL | Result determined to be an outlier at the time of data validation |
UPM | Field duplicate exceeds QC limit with results < 5 x RL. Absolute difference is > RL |
UT | Sample value was blank corrected |
VBB | Sample > 4x spike concentration, flagged by QAO |
VBS | Insufficient sample available to follow standard QC procedures, flagged by QAO |
VBY | Sample received at improper temperature, flagged by QAO |
VBZ | Sample preserved improperly, flagged by QAO |
VBZ15 | Sample not filtered within 15 minutes, flagged by QAO |
VC | Result was calculated by DMT when not provided for DNQ ResQualCode, value 1/2 distance between MDL and RL, flagged by QAO |
VCJ | Analyte concentration is in excess of the instrument calibration; considered estimated, flagged by QAO |
VCO | Sample analyte concentration detectable but <5X blank concentration (contamination), flagged by QAO |
VCQ | Analyte concentration >10X blank concentration |
VCR | Sample analyte concentration >5X blank concentration (True hit), flagged by QAO |
VCS | QC criteria not met due to analyte concentration near RL, flagged by QAO |
VD | EPA Flag - Analytes analyzed at a secondary dilution, flagged by QAO |
VDG | Reporting limits elevated due to sample dilution, flagged by QAO |
VEUM | LCS/LCM recovery not within control limits, flagged by QAO |
VFCL | Field calibration not performed within 24 hours before use, flagged by QAO |
VFDP | Field duplicate RPD above QC limit;flagged by QAO |
VFIF | Instrument/Probe Failure, flagged by QAO |
VFIRL | Invalid RL, suspected laboratory transcription error, flagged by QAO |
VGB | Matrix spike recovery not within control limits, flagged by QAO |
VGBC | CRM recovery not within control limits, flagged by QAO |
VGN | Surrogate recovery not within control limits, flagged by QAO |
VH | Holding time violation occurred, flagged by QAO |
VH24 | Holding time was > 24 hours for Bacteria tests only, flagged by QAO |
VH8 | Holding time was > 8 hrs but < 24 hours for Bacteria tests only, flagged by QAO |
VHB | Result positively biased, flagged by QAO |
VHS | Spike analyte recovery is outside stated control limits, flagged by QAO |
VIL | RPD exceeds control limit, flagged by QAO |
VIP | Analyte detected in field or lab generated blank, flagged by QAO |
VIP5 | Analyte detected in field or lab generated blank, >RL and >1/5 amount detected in associated environmental sample, flagged by QAO |
VIPF | Analyte detected in field blank, data validation code, flagged by QAO |
VIPMDL2 | Analyte detected in field or lab generated blank, >10% lowest detection in batch and >2.2x MDL, flagged by QAO |
VIPMDL3 | Analyte detected in field or lab generated blank, >MDL and >30% lowest detection in batch, flagged by QAO |
VIPRL | Analyte detected in field or lab generated blank, >RL, flagged by QAO |
VIU | Percent Recovery exceeds laboratory control limit, flagged by QAO |
VJ | Estimated value - EPA Flag, flagged by QAO |
VLB | Result negatively biased, flagged by QAO |
VLMQO | Data did not meet laboratory MQO, but met project MQO, flagged by QAO |
VM | A matrix effect is present, flagged by QAO |
VNC | Analyte concentration not certifiable in Certified Reference Material, flagged by QAO |
VNMDL | No Method Detection Limit reported from laboratory, flagged by QAO |
VOL | Result may be biased high due to carry over from prior analysis, flagged by QAO |
VPJM | Result from re-extract/re-anal to confirm original result, flagged by QAO |
VPMQO | Data met laboratory MQO, but did not meet project MQO, flagged by QAO |
VQ | Questionable result, flagged by QAO |
VQAX | When the native sample for the MS/MSD or DUP is not included in the batch reported, flagged by QAO |
VQCA | QA/QC protocols were not met for accuracy, flagged by QAO |
VQCP | QA/QC protocols were not met for precision, flagged by QAO |
VQN | No QA/QC performed or performed but not reported, used with historical data, Batch comment required |
VR | Data rejected - EPA Flag, flagged by QAO |
VRE | Elevated reporting limits due to limited sample volume, flagged by QAO |
VREL | Reporting limits elevated due to change in lab capabilities, flagged by QAO |
VRIL | Data rejected - RPD exceeds control limit, flagged by QAO |
VRIP | Data rejected - Analyte detected in field or lab generated blank, flagged by QAO |
VRIU | Data rejected - Percent Recovery exceeds laboratory control limit, flagged by QAO |
VRVQ | Data rejected - Based on professional judgement QA/QC protocols were not met, flagged by QAO |
VS | Sample Receipt nonconformance, flagged by QAO |
VSCR | Screening level analysis, flagged by QAO |
VTNC | Too numerous to count, flagged by QAO |
VTW | Water quality parameters outside recommended test method ranges, flagged by QAO |
VVQ | Based on professional judgement QA/QC protocols were not met, flagged by QAO |
WE | Sample filtered immediately upon receipt by the laboratory. |
XI | Hydrocarbon response in motor oil range but does not resemble motor oil |